Overview

 

LifeCare PCA™ Infusion System with ICU Medical MedNet™

Product specifications below. Click here for more product information.
Product Numbers List Number20837
Case Information Order Quantity 1 Unit Case Size 1 x 1 Cases per Tier 12 Cases per Pallet 36
Pump Line LifeCare PCA™
Pump Type Pain Management
Dimensions Approximately 8"W 13"H 6"D, excluding pole clamp protrusion and power cord storage.
Weight Approximately 13 lbs. with battery
Warranty The product shall conform to ICU Medical's standard specifications and be free from defects in material and workmanship under normal use and service for a period of one year after purchase. The replaceable battery shall be free from defects in material and workmanship under normal use and service for a period of 90 days after purchase.
Battery Life A fully charged conditioned battery shall provide at least 3 hours of continuous operation from the start of delivery at 5 mL/hr to Dead Battery.
Power Requirements 120 V; 60 Hz; 50 VA max.
Recharge Time After discharge to the LOW BATTERY operative limit, the battery will recharge to at least 80% of its charge capacity in 16 hours, provided it is plugged into rated AC power. The infuser will operate on AC power in the event the battery is open or shorted.
Delivery Rate Accuracy ± 5% at continuous delivery rates greater than 1 mL/hr
Delivery Rates Delivery rate parameters vary depending on mode chosen, drug and concentration selected. Refer to the System Operating Manual for specific details.
PCA Mode: Approximately 1 mL in 35 seconds
Continuous Mode: Variable from 0.1 x concentration (mg/hr or mcg/hr) to 20 x concentration (mg/hr or mcg/hr)
PCA + Continuous Mode: Variable from 0.1 x concentration (mg/hr or mcg/hr) to 20 x concentration (mg/hr or mcg/hr), + PCA dose (mg)
System Alarms Alarms on the LifeCare PCA infuser have three priority levels: high, medium, and low, which occur based on the severity of the alarm condition. Alarms are generated when values for parameters monitored by the infuser move out of a preset range. Empty Syringe, Door Open, Infuser Paused, Low Battery, Dead Battery, Bar Code Not Read, Check Settings, Occlusion, Check Syringe, Check Vial, Check Injector, Pendant Fault, Decimal Point Not Allowed, Check PCA Infuser Set to Relieve Pressure, Malfunction
Occlusion Alarm Pressure 12.5 psi +2.5/-3.5 psi
Lockout Interval Range 5 to 120 minutes in 1-minute increments
Maximum Infusion Pressure 45 psig